3 Operation
7. Empty and clean the reusable secretion canister and the lid according to
section 4.1.3 resp. 4.2.3.
Fig. 15
8. Remove the used bacterial filter before a new patient uses the device or
when it has already been used for one patient for 2 weeks. Move the lever
that is in the horizontal position a quarter turn counterclockwise so that the
lever points vertically downwards subsequently.
EN
9. When the lever is in this position, pull the bacterial filter out of the fitting
and dispose of the bacterial filter properly.
66
Fig. 16
All manuals and user guides at all-guides.com
10. Insert a new bacterial filter (see section 3.2.1).
11. Clean the surface of the device according to section 4.1.2 resp. 4.2.2.
Fig. 17
Fig. 18
Recommendation in the event of an extended storage period: Keep
the device in suitable packaging, e.g. the shipping carton, until it is used
the next time.