4 Maintenance
4.1 Cleaning and care in outpatient and inpatient care
4.1.1 General information
Health risks due to the handling of infectious or
pathogenic germs
Infectious and pathogenic germs in the aspirate cause health risks.
Never operate the device without bacterial filter.
Replace the bacterial filter at least every 2 weeks when used
exclusively for one patient!
Wear suitable disposable gloves in doing so.
The reuse of bacterial filters for multiple patients is prohibited for
hygiene and safety reasons!
For each new patient, processing by
Löwenstein Medical GmbH & Co. KG or by an authorized
service partner of Löwenstein Medical GmbH & Co. KG is strictly
required!
After each aspiration, all components that have come into contact
with the aspirate must be cleaned, disinfected or disposed of.
The aspirate and the parts contaminated with aspirate must be
disposed of properly.
Health risks due to the handling of disinfectants
The use of appropriate protective clothing during disinfection is
recommended.
Follow the disinfectant manufacturer's instructions.
All manuals and user guides at all-guides.com
4.1.2 Cleaning and disinfection of the surface of the device
Repeated cleaning and disinfection procedures may result in minor discolora-
tions of the plastic components of the housing. However these do not affect
the function of the device.
Damage of the device due to improper cleaning agents
Do not use disinfectants containing acetone. These may damage or
disfigure the housing parts as well as the secretion canister lid.
Follow the instructions for use provided by the manufacturers of
the used disinfectants, particularly concerning the compatibility with
materials and surfaces as well as the concentration information.
Löwenstein Medical GmbH & Co. KG recommends
"Sekusept
aktiv" for immersion disinfection of the accessories
®
and "Incidin
Plus" or "Incidin
®
the device.
Clean the surfaces of the device regularly and disinfect them at least
weekly.
The device can be wiped with a damp, lint-free cloth.
Follow the instructions in section 4.1.1 for wipe disinfection.
If the interior of the device comes into direct contact with liquids or
solids, the device must be inspected by Löwenstein Medical GmbH
& Co. KG or by a service partner authorized by Löwenstein Medical
GmbH & Co. KG.
Liquid" for wipe disinfection of
®
EN
67