3 Operation
Hazard of persons due to improper handling
•
Please read through sections 3.1 and 3.2 carefully!
•
Only perform aspiration in the respiratory tract after instruction by
specialized personnel!
•
Only use suitable sterile disposable catheters for aspiration!
Commercially available sterile disposable catheters from CH10 to
CH16 that are intended for the use in the tracheal area can be
connected.
•
If aspiration procedures are carried out too often, slight bleeding
may occur.
•
When using the device for the treatment of children, the vacuum
setting "Low" must be used.
Malfunction due to aspirated secretion
•
Change the reusable secretion canister when it is half-full to curtail
frothing. If the reusable secretion canister is full, the mechanical
overflow protection is triggered. This disrupts the aspiration
process.
•
Switch off the device when emptying the reusable secretion
canister.
•
If secretion has been aspirated into the device, it must be properly
processed by Löwenstein Medical GmbH & Co. KG or by a service
partner authorized by Löwenstein Medical GmbH & Co. KG!
Damage of the device due to insufficient
acclimatization
•
After the device has been exposed to temperatures between
-25°C and +70°C during transport/storage according to the tech-
nical data (see section 7), it must firstly acclimatize for ca. 2 h at
room temperature (ca. 20°C) before the intended use is possible.
EN
57