3 Operation
Hazard of persons due to improper handling
Please read through sections 3.1 and 3.2 carefully!
Only perform aspiration in the respiratory tract after instruction by
specialized personnel!
Only use suitable sterile disposable catheters for aspiration!
Commercially available sterile disposable catheters from CH10 to
CH16 that are intended for the use in the tracheal area can be
connected.
If aspiration procedures are carried out too often, slight bleeding
may occur.
When using the device for the treatment of children, the vacuum
setting "Low" must be used.
Malfunction due to aspirated secretion
Change the reusable secretion canister when it is half-full to curtail
frothing. If the reusable secretion canister is full, the mechanical
overflow protection is triggered. This disrupts the aspiration
process.
Switch off the device when emptying the reusable secretion
canister.
If secretion has been aspirated into the device, it must be properly
processed by Löwenstein Medical GmbH & Co. KG or by a service
partner authorized by Löwenstein Medical GmbH & Co. KG!
Damage of the device due to insufficient
acclimatization
After the device has been exposed to temperatures between
-25°C and +70°C during transport/storage according to the tech-
nical data (see section 7), it must firstly acclimatize for ca. 2 h at
room temperature (ca. 20°C) before the intended use is possible.
EN
57