1 User information
1.6 Basic safety instructions - ATTENTION!
Warning of damage due to improper power supply
Improper operation causes overvoltage in the device, which may be transmitted
to the operator.
Ensure prior to startup that the mains supply for connecting the
secretion aspirator Allegra M30 is suitable at 100 V to 240 V AC
and a mains frequency of 50-60 Hz.
Ensure prior to startup in UL listed markets such as the USA and Canada
that the mains supply is designed for a supply voltage of 120 V AC.
Only use the supplied power supply unit with the secretion
aspirator Allegra M30 (type: GTM91099-6015-3.0-T2
from GlobTek Inc.).
Risk of application under conditions that are not approved
The device is not intended to be used in medical rooms where potential
equalization is required (e.g. heart surgery).
The device is not intended to be used in hazardous areas/in the MRI
environment.
The device is intended for mobile, temporary outdoor use (in accordance
EN
with IP protection class 22), but not for permanent outdoor location and
operation.
50
Health risks due to the handling of infectious or pathogenic
germs
Infectious and pathogenic germs in the aspirate cause health risks.
Never operate the device without bacterial filter.
Prior to each application, make sure that the bacterial filter is dry and clean
to ensure faultless operation. Pay attention that the bacterial filter does not
get moist during the performance of cleaning/disinfection.
Always perform the aspiration using a suitable sterile disposable catheter.
The suction tube must never come into direct contact with the aspiration
site.
Follow the hygiene, cleaning and decontamination instructions.
All manuals and user guides at all-guides.com
Hazard of persons due to strangulation
Risk of damage due to electromagnetic phenomena
Medical electrical devices are subject to special precautionary measures regard-
ing electromagnetic compatibility and must be installed and operated in accord-
ance with the EMC information provided in the accompanying documents (see
section 8).
Hazard of persons due to improper handling
People may strangle themselves on the tubing or the mains cable, especially
if tubes resp. cables are unduly long.
Ensure that no unauthorized/uninvolved persons are near the device during
aspiration.
Store the device incl. accessories in the shipping carton until the next time
of use.
Only use the device for its intended purpose.
Never use the device for aspiration in the low vacuum range
(e.g. thoracic drainage).
If aspiration procedures are carried out too often, slight bleeding may occur.
When using the device for the treatment of children, the vacuum setting
"Low" must be used.
When using the power supply unit, ensure that it is connected to the
aspirator firstly and afterwards to the mains supply (100 V to 240 V AC).
The disconnection of the power supply unit from the mains supply must be
performed in exactly the opposite sequence (first remove the power supply
unit from the mains supply (100 V to 240 V AC) and then remove it from
the aspirator).
Never touch parts of other devices than ME devices in the patient environ-
ment and the patient simultaneously.