4.2 Cleaning and care in the homecare area
4.2.1 General information
Health risks due to the handling of infectious or
pathogenic germs
Infectious and pathogenic germs in the aspirate cause health risks.
Never operate the device without bacterial filter.
Replace the bacterial filter at least every 2 weeks when used
exclusively for one patient!
Wear suitable disposable gloves in doing so.
The reuse of bacterial filters for multiple patients is prohibited for
hygiene and safety reasons!
Prior to passing the device on to another patient resp. to other
persons, processing by Löwenstein Medical GmbH & Co. KG or by
an authorized service partner of Löwenstein Medical GmbH & Co.
KG is strictly required! For that purpose observe the information
regarding the further use of the device (see section 4.3).
After each aspiration, all components that have come into contact
with the aspirate must be cleaned, disinfected or disposed of.
The aspirate and the parts contaminated with aspirate must be
disposed of properly.
Health risks resp. damages of the device due to the
handling of disinfectants
If you have the possibility to use disinfectants, please observe the
information in section 4.1. Otherwise please follow the indications
of section 4.2.2 till 4.2.5!
All manuals and user guides at all-guides.com
4.2.2 Cleaning of the surface of the device
The device can be wiped with a damp, lint-free cloth. Repeated cleaning
procedures may result in minor discolorations of the plastic components of the
housing. However these do not affect the function of the device.
Clean the surfaces of the device regularly.
If the interior of the device comes into direct contact with liquids
or solids, the device must be inspected by
Löwenstein Medical GmbH & Co. KG or by a service partner
authorized by Löwenstein Medical GmbH & Co. KG.
4 Maintenance
EN
69