3 Operation
3.3.2 Setting the vacuum
1. Switch on the device by pressing the on/off button (fig. 11(F)).
2. Select the desired vacuum range by pressing the corresponding button
("Low", "Medium" or "High") on the control panel (fig. 11(A, B, C)).
As described in section 3.3.1, the vacuum setting buttons correspond
to certain vacuum ranges.
3. You can perform fine adjustments of the set vacuum via the additional air
opening on the fingertip (fig. 12).
EN
64
Fig. 12
4. Switch off the device by holding the on/off button (fig. 11(F)) pressed for
1-2 seconds.
All manuals and user guides at all-guides.com
3.3.3 Aspiration
1. Perform the aspiration procedure as you were instructed by specialized
2. Control the aspiration procedure and the desired vacuum via the additional
3. Decommission the device after the aspiration procedure according to
personnel.
air opening on the fingertip (fig. 12).
section 3.4.
In any case, please take note of section 2.4 and the important inform-
ation on the handling of the bacterial filter! Ensure that the battery has
a sufficient capacity during the aspiration procedure. The battery status
display must signalize green or orange. If the display turns red, the
power supply unit must be plugged in immediately. Only in this case the
aspiration can be continued without any interruption.