4.2.5 Hygienic cleaning and disinfection of the rinsing bottle
1. Remove the lid of the rinsing bottle and rinse both parts under running
water.
2. Place the rinsing bottle and the lid in a water bath at 65°C temperature
for 15 minutes once a day.
3. Afterwards, rinse the components thoroughly and let them dry.
4. Place the lid back on the rinsing bottle.
Löwenstein Medical GmbH & Co. KG recommends to replace the rinsing bottle
every 4 weeks if it is used and disinfected frequently!
4.3 Further use of the device
The secretion aspirator Allegra M30 is suitable for further use.
However, prior to passing the device on to other patients resp. persons, it has
to be processed professionally. For this purpose, please hand the secretion
aspirator Allegra M30 over to Löwenstein Medical GmbH & Co. KG or to
qualified personnel authorized by Löwenstein Medical GmbH & Co. KG. In this
regard, please observe the indications in section 6.1!
All manuals and user guides at all-guides.com
4.4 Maintenance and service
Risk of impairments of performance due to aging
effects
Only use the device within its indicated service life of 5 years. If you
notice impaired performance or other problems, follow the instructions
on troubleshooting in section 5 and contact the service if necessary.
Risk due to maintenance and service during operation
Maintenance and service must not be performed while the device is in
use! The device must be switched off before maintenance and service
work are performed.
Risk due to the performance of maintenance and
service by unauthorized persons
Maintenance and service may only be performed by specialized
personnel authorized by Löwenstein Medical GmbH & Co. KG
or by Löwenstein Medical GmbH & Co. KG itself.
Risk due to the performance of maintenance and
service without sufficient documents
Maintenance and service may only be performed considering the
service instructions for the secretion aspirator Allegra M30.
Löwenstein Medical GmbH & Co. KG will also provide additional docu-
ments on request, in case these can support the service staff.
WARNING: Modifications of the ME device are not
permitted.
4 Maintenance
EN
71