WARNING
There is a risk of an electric shock if the foot pedal or the ON/OFF key of
the motor is pressed on the user interface of the SIROEndo and the mi-
cromotor is not connected to the SIROEndo. In this case, the "Motor"
LED lights up for one second.
The Rpm display indicates the actual speed of the contra-angle hand-
piece if the appropriate reduction ratio is selected. Check whether the re-
duction value on the display agrees with the reduction ratio of the contra-
angle handpiece being used.
In case of noticeable malfunctions, stop working with the instrument im-
mediately and contact your service engineer.
Do not apply force to rotate the swivel arm of the unit more than 180° in
both directions.
Never under any circumstances insert the SIROEndo software stick in a
USB port of a PC.
Never insert any USB device other than the designated SIROEndo mem-
ory stick in the USB port of the SIROEndo.
Never under any circumstances connect the SIROEndo unit to a PC via a
USB cable.