Sterilization
Motor and contra-angle handpiece
Power supply
Side effects
Disposal
60 78 104 D 3488
D 3488.201.01.07.02 11.2009
All manuals and user guides at all-guides.com
The micromotor must be uncoupled from the motor cable prior to steril-
ization. The motor cable itself is not suitable for sterilization.
Maximum sterilization temperature for motor without motor cable:
134 °C (273.2 °F).
You must remove the contra-angle handpiece before sterilizing the mo-
tor.
The RPM display indicates the effective rpm of the instrument if the suit-
able reduction ratio is selected. Check to make sure that the reduction ra-
tio displayed is identical to the reduction ratio of the contra-angle
handpiece attached to the motor.
In case of noticeable malfunctions, stop working with the instrument im-
mediately and contact your service engineer.
Excessive bending or kinking of the micromotor cable should be avoided
to prevent damage.
The SIROEndo is operated with a power supply of 100 V - 240 V, 47 - 63 Hz.
As the manufacturer, Sirona Dental Systems GmbH provides on written
request qualified technical personnel, wiring diagrams, parts lists, and
other useful information for the repair and maintenance of those parts
which the manufacturer considers to be repairable.
Until now, no side effects (e.g. necrosis) have been determined in connection
with clinical use of the Sirona SIROEndo.
It generally applies that any disposal of this product must comply with the rel-
evant national regulations.
Please observe the regulations applicable in your country.
Within the European Economic Community, Council Directive 2002/96/EU
(WEEE) requires environmentally sound recycling/disposal of electrical and
electronic devices.
Your product is marked with the adjacent symbol. Disposal of your product
with domestic refuse is not compatible with the objectives of environmentally
sound recycling/disposal.
The black bar underneath the "garbage can" symbol means that it was put
into circulation after Aug. 13, 2005. (see EN 50419:2005)
Please note that this product is subject to Council Directive 2002/96/EU
(WEEE) and the applicable national law of your country and must be recycled
or disposed of in an environmentally sound manner.
Please contact your dealer if final disposal of your product is required.
1 Warning and safety information
5