12.9
Vacuum pump
12.9.1
Changing vacuum pump oil
DANGER – Risk of electric shock from
dangerous voltage!
Electric shocks resulting from contact with
live parts (e.g. master switch, printed cir-
cuit boards) can cause injury, heart failure
and death.
¶
Disconnect the mains plug before open-
ing the ACS 652.
ATTENTION – Risk of burns from hot
surfaces
Contact with the hot vacuum pump surface
will cause severe burns.
¶
Allow the vacuum pump to cool down.
¶
Wear protective gloves.
i
The vacuum pump oil must always be changed after
10 hours of operation. The prompt "
appears on the screen when the vacuum pump oil
needs changing.
i
Use the vacuum pump oil (item number
SP00100086) as specified by Bosch.
2
Fig. 8:
Vacuum pump
1 Sight class
2 Oil fill plug
3 Drain plug
|
S P00 D00 158
2019-02-21
REPLACE OIL
"
3
1
!
Do not use excessive force when securing and re-
moving the oil drain and oil filling screw.
i
The container for catching oil should have a capacity
of approx. 1 l.
1. Place a vessel under the drain.
2. Remove drain plug and oil fill plug from vacuum
pump.
3. Drain off all the oil.
4. Screw on drain plug finger-tight.
5. Fill in vacuum pump oil.
6. Start the vacuum phase.
7. Check the oil level (sight glass).
i
The oil level should be in the middle between the
"full" and "empty" marks.
12.9.2
Reset of oil change interval
!
Observe the ACS 652 menu prompting.
1. Call up start screen.
2. Press <3> and <Flush> simultaneously and release.
The message
ENTER CODE 0000
5555
and confirm with E.
3. Enter code
4. Select "CHANGE OIL" with o u and confirm with E.
Current runtime of vacuum pump is displayed.
5. Press <MENU> until vacuum pump runtime is set to
zero.
"
Oil change interval is reset.
i
Return with <Stop>.
is displayed.
Robert Bosch GmbH