Pfeiffer Vacuum CENTERONE Instructions D'installation page 12

Masquer les pouces Voir aussi pour CENTERONE:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 14
Installation
B-type USB
B-type USB
4
3
1
2
A-type USB
A-type USB
1 2 3 4
12
Pin
Signal
Infeed for relays with a higher switching capacity
25
+24 V DC, 200 mA
1, 7
GND
2
n.c.
1)
200 mA fuse protection with a PTC element, self-resetting after the CenterTwo or CenterThree is
switched off, or the RELAY connector is removed. Meets the requirements for grounded protective
extra-low voltage.
Fig. 14: Sub-D socket
The B-type USB port facilitates direct communication with the CenterOne/CenterTwo/
CenterThree via a computer (e.g. firmware update, parameter storage (read/write)).
Pin
Assignment
1
VBUS (+5 V)
2
D-
3
D+
4
GND
 Connect a USB port with a shielded cable (electromagnetic compatibility) to the USB
terminal on the back of the device.
USB Type B
The A-type USB port, which has master functionality, is located on the front and is used
for the connection of a USB memory stick (for instance, for firmware updates, parameter
storage (read/write), data logger).
Pin
Assignment
1
VBUS (+5 V)
2
D-
3
D+
4
GND
 Connect the USB memory stick to the USB port on the front of the device.
1)