CONTROL terminal
(CenterOne)
Pin
Signal
3
Status
4
HV_H / HV_EMI
5
Infeed ground (GND)
6
n.c.
7
Degas
8
Infeed (+24 V DC)
9
n.c.
10
Identification
11
Infeed (+24 V DC)
12
Signal ground
13
RxD
14
TxD
15
Housing
Fig. 11: Sub-D socket
This terminal can be used to:
● read out the measuring signal,
● perform an isolated evaluation of the state of the switching function and error monitor-
ing,
● switch the high-vacuum measuring circuit on and off (only with the cold-cathode trans-
mitter PTR 225).
Dangerous contact voltage
Voltages above 30 V AC or 60 V DC are deemed to be dangerous in the event of contact
in accordance with EN 61010.
 Only apply a grounded protective extra-low voltage (PELV).
 Connect peripheral components with a shielded connecting cable (electromagnetic
compatibility) to the CONTROL terminal on the back of the device.
Pin assignment and contact positions
The pins on the 15-pin sub-D connector are assigned as follows:
Pin
Signal
1
Analog output -5 ... +13 V DC
2
Analog output GND
Switch function 1
3
4
5
6
HV_V on (+ 24 V)
HV_H off (0 V)
7
+24 V DC, 200 mA
8
GND = housing
DANGER
2)
Pressure higher than
threshold or device
switched off
1)
1)
Installation
Pressure lower than
threshold
9