Télécharger Imprimer la page

PROTEOR 4 AXIS Notice D'utilisation page 13

Genou

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 1
D. Conversion to free-moving knee (Version V)
This operation is irreversible.
E. Finishing
The socket must be in contact with the tube holder or its collar at maximum bend. Any
other contact zone is prohibited.
For the locking version, check that the lock is not acting as a stop.
F. Installing the lock control handle 1X110
For 1M102V(-P6) only
This handle is supplied in its "automatic holding in unlocked position" version. To obtain a handle with automatic return, please see
below.
Maximum unlocking stroke in the "automatic holding in unlocked position" version. 14 mm.
Maximum unlocking stroke in the "automatic return" version: 15 mm.
For certain operations described below, it may be necessary to dismantle the handle mechanism.
In this case the mechanism dismantling and reassembly operations must be performed as set out below:
Disassemble the handle by unscrewing, in order, screws 8, 9 and 7 (to be able to access screw 8, the handle must be in the
down position, with hole 4a in this position enabling you to insert a 2 mm Allen key)
To reassemble the mechanism, follow the same procedure, refitting, in order, screws 8, 7 and then 9.
For any operations with the handle, the tab 3 must be in the up position. If it is in the down position, screw 8 may rise back into the
mechanism and become inaccessible. Should this occur, dismantle the mechanism, reinsert screw 8, and then refit as described
above.
Modification to obtain automatic handle return:
Disassemble the handle as set out above
Using a screwdriver, lift the tab 3a
Shave the hooks 3b
Use a cutter to eliminate any burrs that could impede
performance
Reassemble the mechanism.