Menu description
10.4.4
Calibration / Manual
If the calibration value to be entered is known, a "dry calibration" (immediate entry of the
value without a previous calibration process) can be conducted.
However, this method is not very accurate as no consideration is given to the on-site
conditions (back-pressure, viscosity, cable cross-chapters and lengths etc).
Fig. 82: Calibration / Manual
10.4.4.1 Calibration data table
The pump capacity in l/h is allocated to the pump's electronics on the basis of the
calibration values entered in the table.
Example: A specified value of 23.3 ml results in a pump capacity of 1.4 l/h.
NOTICE!
These values relate to the metering medium of water at 20°C.
Pump
Calibration / Manual
417101442 Rev. 12-02.2019
Pump capacity [l/h]h]
at 50 (60) Hz
0,2 (0,24)
0,5 (0,6)
0,9 (1,1)
1,4 (1,7)
104
Pump calibration
value [ml]
at 50 (60) Hz
3,33 (4,00)
8,33 (10,0)
15,0 (18,3)
23,3 (28,3)