Overview Of Maintenance And Care For 1057 - KaVo PROMATIC 1057 Serie Instructions De Service

Table des Matières

Publicité

A 8.9
What
Strainer in spittoon bowl
Suction hoses
Water-operated
saliva ejector
Strainers in suction
connector
Instruments
INTRA K-Lux motor 192
INTRAmatic motor 181 L
INTRAmatic motor 181 M After each treatment
Heavy particle collector
Paintwork surfaces
Upholstery
______________________________________________________________________________________________________________________
Spittoon bowl
KAVOLUX 1410
Amalgam separator
After each treatment
Multi System Type 1
KaVo PROMATIC 1057
O
VERVIEW OF MAINTENANCE AND CARE FOR
When
How
serveral times a day
Clean strainer under running water.
Dispose of contents properly.
After each treatment
Aspirate a tumbler full of cold water
Do not aspirate foaming cleansers.
In the evening
Disinfect suction hoses with KaVo Aseptoclean 2.
Observe instructions included in Aseptoclean 2 package.
Several times a day
Actuate backflush.
After surgical work,
Check strainers. Replace as required, at the latest when suction
otherwise daily
decreases. Cleaning cannot be recommended for reasons of
hygiene.
After each treatment
Change, clean, lubricate and sterilize or disinfect outer surfaces.
2122 ROTA Spray - order no. 411 9850
2142 ROTA Spray 2 - order no. 411 7510
KaVo Spray - order no. 411 9830
At noon and in the
evening
Lubricate. Allow instruments to dry well on instrument stand..
After each treatment
Disinfect exterior.
After each treatment
Disinfect exterior.
At noon and in the
Lubricate.
evening
Disinfect exterior.
At noon and in the
Lubricate.
evening
As required
Sterilize
Weekly or as required
Open heavy particle collector with tool (order no 411 1843, from
accessories) and clean. Dispose properly of amalgam. Drain water
before reinstallation even if bowl does not need to be cleaned to
prevent overflow when opening the next time.
Only hand-tighten bowl (do not use tool).
Monthly or
Wipe surfaces with KaVo paintwork preserver (order no. 411 1843)
as required
and a soft cloth. Do not use abrasive cleansers or cleansers con-
taining acid. See sections A 1.3-1.4 regarding care and disinfection.
Monthly or
Use KaVo Elastoclean spray II (order no. 489 3467 0.5 litre or
as required
order no. 489 3469 1.0 litre) and wipe upholstery off with a
soft cloth See section A 1.3 - 1.4 regarding care and disinfection
Daily or
Clean spittoon bowl according to instructions supplied with
as required
Aseptoclean package.
As required
Wipe the reflector with water and a soft cloth when the reflector is
cold. Do not use abrasive cleansers or cleansers containing acid.
Do not spray disinfect.
Aspirate 1 tumbler (approx. 100 ml) of cold water into suction hoses
used. Rinse spittoon bowl.
Daily
Check filter drawer and clean as necessary.
In the evening
Use Aseptoclean
Aseptoclean package.
GB
1057
®
2 according to instructions supplied with
DE
B
ENUTZERHINWEISE
Description
in OMA instr.
Section A 8.6.5
Section A 8.6.1
Section A 8.6.2
Section A 8.6.4
Note OMA
instructions of
instruments
Note OMA
instructions of
motors
Section A 8.6.7
Note OMA
instructions for
KAVOLUX 1410
Note OMA
instructions for
Multi System
Type 1 amalgam
separator
83