ECHNICAL DATA AND
S
ERVICIO DEL ELEMENTO DEL LA
REGQUIREMENTS
AYUDANTE
Drilling template No.
1057 S, T
1057 G
Installation plan No.
1057 T
1057 S
1057 G
Electric supply cable
Free end above floor
Input voltage
Frequency
Factory-set input voltage
Max. power consumption
at 100 - 120 V
at 220 - 240 V
Fuse (supplied by customer)
Potential equilibration line
Above floor
Heat output
Approval symbols
Water
According to local waterboard regulations
Water quality
Water hardness
Total hardness:
pH-value
Water supply filtration
Water connection
above floor
Areas with brackish stagnant water should be excluded.
Water inlet pressure
Water inlet flow rate
C
OMMANDE ENSEMBLE ASSISTANTE
485 1193/001
486 7403
486 7419 Sheet 2
486 7419 Sheet 3
486 7419 Sheet 4
3 x 1.5 mm
1000 mm
110/110/120 V AC
130/220/230/240 V AC
50/60 Hz
See the name plate
500 VA
900 VA
depending on mode
Automatic C16 or
screw-type fuse 10 A
See VDE 0107
1000 mm
72 - 3000 kJ/h
depending on mode
CE
Drinking water
1.5 - 2.14 mmol/l
^ = 8,4 - 12 dH
7,2 - 7,8
80 µm
R 1/2
40 mm
2.5 - 6.0 bar
wihtout jet pump 5 l/min
with jet pump 8 l/min
depending on water pressure
FR
Drain connection
above floor
Drainage rate
Gradient of water drain pipe
2
Air connection
above floor
Air inlet pressure
Air inlet flow rate
Suction connection
above floor
Suction vacuum / Suction vacuum rate
≥ 65 mbar / ≥ 500 Nl/min at inlet into unit
≥ 19 mbar / ≥ 280 Nl/min at suction nozzle
Values applicable for KaVo measuring set, Order No. 411 8500
Weight
For further information on the packages, see assembly
instructions, Section B3.
Floor characteristics: The quality of the floor finish must
correspond to the load carrying capacity for buildings
according to DIN 1055 sheet 3; compressive strength
according to DIN 18 560, part 1.
1)
1)
DIN 1988 as well as twin 6 (issue 01/94) are in force. A safety
device type BA (three chamber system) is to be installed
before each dental unit.
2)
In the event of high water hardness, a water softening unit
2)
must be installed in the ion exchange process with disinfection
system. An excessively low water hardness may promote
algal growth.
O
'
PERATING THE ASSISTANT
S UNIT
GB
Ø 40 mm
wihtout jet pump 5 l/min
with jet pump 8 l/min
from unit at least
10 mm per meter
5.2 - 7 bar
max. 80 l (S.T.P.)/min
Ø 40 mm
static < 180 mbar
gross/net
- / approx. 190 kg
KaVo PROMATIC 1057
20 mm
R 1/2
40 mm
20 mm