G.2 "Transmission" function
Maximum rated voltage in common mode: 60 VDC or VAC.
G.2.1 DC voltage
Range
Measurement
0/10V
0 V/+12 V
15V
0 V/+15 V
•
Temperature coefficient < 15 ppm/°C from 0°C to 18°C and from 28°C
to 50°C.
•
Rise time: < 1ms (0V to 15V across 1-M
•
Internal resistance: ≤ 1 Ω.
•
VLF noise: < 1mV (at F< 100Hz).
G.2.2 Direct current
Range
Measurement
0/20mA
+500 A/+24 mA
4/20mA
+3.2 mA/+24 mA
25mA
+500 A/+25 mA
•
Temperature coefficient < 20 ppm/°C from 0°C to 18°C and from 28°C
to 50°C.
•
Rise time: < 500 µS (0 to 20 mA across a 20- load).
•
VLF noise: < 1µA (at F< 100Hz).
G.3 "Continuity" function
Maximum rated voltage in common mode: 60 VDC or VAC.
Range
Resolution
Continuity
Open/Closed
Minimum
range
resolution
Minimum
range
resolution
Considered closed at Z=[0...11KΩ (+/-10%)]
and open at Z=]11KΩ (+/-10%)...∞[
I < 5mA
Precision/1 year
1 mV
0.015% L + 2 mV
1 mV
0.015% L + 2 mV
load).
Precision / 1 year
1 A
0.015% L + 2 A
1 A
0.015% L + 2 A
1 A
0.015% L + 2 A
Remarks
45/47