Note 1:
When the range is exceeded (limits determined in the next chapter), the
following pictogram
indicates:
- :----
if the measured value is below the low limit.
+ :----
if the measured value is above the high limit.

B.1.5 Starting (after powering on)

At powering on (insertion of the AA batteries or battery pack), the instrument
automatically starts functioning (loading of the software into memory). It is
recommended not to connect the instrument to any external circuits during
this period.
To avoid any signal conflicts, the instrument then enters measurement mode.

B.1.6 Operating modes

There are 5 main operating modes:
Voltage measurement,
Current measurement,
Voltage transmission,
Current transmission,
Continuity test.
The functional and electrical characteristics not to be exceeded are indicated
below:
B.1.6.1
DC voltage measurement
The following ranges are available:
Range
Resolution
1 mV (or 10 mV or
(display)
100 mV)
Extent of
-2 V to +12 V
the range
appears and the measurement display (5th field)
0/10 V
(Process)
1 mV (or 10 mV or 100
mV)
-2 V to +25 V
25 V
1 mV (or 10 mV or
100 mV)
-5 V to 50 V
50 V
17/47