Cyclic ramp signal configuration?
The figure below shows the type of cyclic ramp signal that can be generated
along with its parameters:
Programming parameters in Cyclic ramp mode
NH
NB
T0
Tb
Td
To : The moment the start of generation key is pressed
Td : Delay
Tb : Low Level Duration
Tm : Rise (rise time)
Th : High Level Duration
Tf : Fall (fall time)
NB : Low Level
NH : High Level
The LOW Level and HIGH level are given:
As a percentage of the range if scaling mode is ON.
In volts or mA if the scaling mode is OFF and depending on the type
of quantity transmitted (voltage or current transmission).
The Low Level Duration, Rise, High Level Duration, Fall and Delay
times are given in seconds and are limited to a maximum of 1000s.
The Repetitions field gives the number of ramp signals to be generated.
There can be a maximum of 1000 repetitions.
34/47
P
ROGRAMMING THE MODES
Th
Tf
Tm