Measurement mode provides for display of the Minimum (bottom left), Mean
(bottom centre) and Maximum (bottom right) values since the last Min/Max
Reset command.
This command is accessed through key F2.
Using the navigation key, go onto the Min/Max Reset field
Press VAL to confirm.

C.1.3 Current measurement / auxiliary functions

Auxiliary functions can be defined for all the ranges. Examples include:
Supply of a +24 voltage to the sensors or not
mode
Linear or quadratic scale (for process ranges only).
Hart compatibility (for process ranges only).
After selecting the Function and Range, the auxiliary functions described
above must be defined.
passive or active mode?
Using key F1, go onto the Loop Supply field.
Using the navigation keys (
(Passive or Active).
This active or passive mode also concerns the Hart mode.
Scale?
Using key F1, go onto the Scale field.
Using the navigation keys (
mode.
When quadratic scaling is activated, the CP6632 takes the square root of its
input and displays the result as a percentage. For example, if the CP6632 is
connected to the output of a differential pressure transmitter it displays a
result proportional to the flow.
If the input current x varies between a and b, scaling complies with the
formula below:
The scaling curve for the 0-20mA range is as follows:
24/47
P
ROGRAMMING THE MODES
↑ ↑ ↑ ↑
and
↑ ↑ ↑ ↑
and
(
=
+
y
a
b
a
Passive or active
↓ ↓ ↓ ↓
), select the mode ON or OFF
↓ ↓ ↓ ↓
), select Linear or Quadratic
) (
)
x
a
(
)
b
a