CP6632
G. T
ECHNICAL SPECIFICATIONS
The indicated levels of precision apply at temperatures of +18°C to +28°C,
unless otherwise specified, and are expressed in ± (n % R + C) where R =
Reading and C = Constant expressed in practical units and given for a
confidence interval of 95%.
They apply to an instrument placed in the reference conditions defined
elsewhere, after fifteen minutes of warming up.
The precision includes the precision of the reference standards, the non-
linearity, the hysteresis, the repeatability and the long-term stability over the
period mentioned.
G.1 Measurement function
Maximum rated voltage in common mode: 60 VDC or VAC.
G.1.1 DC voltage
Range
Measurement
0/10 V
-2 V/+12 V
25 V
-2 V/+25 V
50 V
-5 V/+50 V
•
Temperature coefficient < 15 ppm L /°C from 0°C to 18°C and from
28°C to 50°C.
•
Rin = 1 M Ω +/− 1%
•
Common mode rejection: ≥ 120 dB at 50 and 60 Hz.
G.1.2 Direct current
Range
Measurement
0/20 mA
-6 mA/+24 mA
4/20 mA
3.2 mA/+24 mA
25 mA
-6 mA/+25 mA
•
Temperature coefficient < 20 ppm/°C from 0°C to 18°C and from 28°C
to 50°C.
•
Possible loop power supply = 24 V ± 10%.
•
Rin < 30 Ω
•
HART compatibility: R = 250 Ω +/− 5%
•
Common mode
44/47
T
ECHNICAL SPECIFICATIONS
Minimum
range
resolution
1 mV
1 mV
1 mV
Minimum
range
resolution
1 µA
1 µA
1 µA
≥ 120 dB at 50 and 60 Hz.
rejection:
Precision / 1 year
0.015% L + 2 mV
0.015% L + 2 mV
0.015% L + 4 mV
Precision / 1 year
(CI : 95%)
0,015% L + 2 µA
0,015% L + 2 µA
0,015% L + 2 µA