G. TECHNICAL SPECIFICATIONS
The precision expressions mentioned herein apply from + 18°C to + 28°C, unless otherwise specified, and are expressed
in ± (n % L + C) where R = Reading and C = Constant expressed in practical units, for a confidence interval of 95%.
They apply to a device positioned in the reference conditions defined below:
Prior power-up of the device for temperature adjustment for 10 minutes
Following a significant thermal shock, it is advisable to allow the device to stabilize its temperature to use the
internal reference junction (CSF) with the utmost precision.
The precision includes the precision of the reference calibrations, the non-linearity, hysteresis, repetitiveness and long-
term stability over the time period mentioned.
G.1
Thermocouple measurement function
Rated maximum voltage in common mode: 60 VDC or VAC.
G.1.1
Constant voltage (TC6621,TM6602 and TM6630)
The voltage measurement is made by configuring the device as follows:
Thermocouple: indifferent.
Unit: mV.
CSF: OFF.
Gauge
Scope of measurement
100 mV
-10 mV to 100mV
Temperature coefficient < 10 ppm R /°C from 0°C to 18°C and from 28°C to 50 °C.
Rin = 1 MΩ ±1%
G.1.2
Température per thermocouples (TC6621,TM6602 et TM6630).
Type of sensors:
- Standardised in accordance with CEI 584-1/1995 (Thermocouples K, T, J, E, S, B, N)
- In accordance with Din 43710 (thermocouples U and L)
- In accordance with the HOSKINS table (thermocouple C)
- In accordance with the ENGELHARD table (platinum thermocouple).
- Standardised in accordance with ASTM E 988-96 /2002 (Couple D W3Re/W25Re)
- Standardised in accordance with ASTM E 1751-00 /2000 (Couple G W/W26Re)
Thermocouple Measurement function
Resolution (min)
Precision / 1 year
1 µV
0.020% R + 3 µV
G
NTA47230-300A6
93 / 101