CP6632
Single ramp signal configuration?
The figure below shows the type of single ramp that can be generated along
with its parameters:
Programming parameters in Single ramp mode
T0
Tdt
To : The moment the start of generation key is pressed
Td : Delay
Tdt : Total Duration
NB : Low Level
NH : High Level
The LOW Level and HIGH level are given:
•
As a percentage of the range if scaling mode is ON.
•
In volts or mA if the scaling mode is OFF and depending on the type
of quantity transmitted (voltage or current transmission).
The Total Duration corresponds to the incrementation time it takes to go
from Low Level to High Level (and vice versa with decrementation). It is
given in seconds and is limited to a maximum of 1000s.
The Delay corresponds to the timeout that can be programmed between the
moment the start of transmission key is pressed and the actual start of
generation. It is given in seconds and is limited to a maximum of 1000s.
32/47
P
ROGRAMMING THE MODES
NH
NB
Td
T0
Tdt
Td
T0
Tdt