C.1.6 Continuity

CONTINUITY mode is a MEASUREMENT mode applied to the current channel
(I+, I- terminals).
It indicates whether the circuit (the loop) is open or closed. The circuit must
be resistive with no active elements. The loop shall be considered closed at
loop impedances of 0 to 11KΩ (+/-10%), and open at impedances above
11KΩ(+/-10%).
Select the measurement mode using key F2 (mode menu).
Using the navigation keys, move down through the menu to the
Measurement field.
Press the VAL key to confirm.
The type of function (voltage or current or Continuity) is selected
using key F1 (configuration menu).
Using the navigation key, go onto the Function field and press VAL
to confirm.
Using the navigation key (↑ ↑ ↑ ↑ and ↓ ↓ ↓ ↓ ), select the Continuity function
and press VAL to confirm. It is possible to enter the Function field
using key F2 and to select the function with the navigation keys (↑ ↑ ↑ ↑
and ↓ ↓ ↓ ↓ ).
Confirming the Continuity function calls up the following screen: