CP6632
C.1.6 Continuity
CONTINUITY mode is a MEASUREMENT mode applied to the current channel
(I+, I- terminals).
It indicates whether the circuit (the loop) is open or closed. The circuit must
be resistive with no active elements. The loop shall be considered closed at
loop impedances of 0 to 11KΩ (+/-10%), and open at impedances above
11KΩ(+/-10%).
•
Select the measurement mode using key F2 (mode menu).
•
Using the navigation keys, move down through the menu to the
Measurement field.
•
Press the VAL key to confirm.
•
The type of function (voltage or current or Continuity) is selected
using key F1 (configuration menu).
•
Using the navigation key, go onto the Function field and press VAL
to confirm.
•
Using the navigation key (↑ ↑ ↑ ↑ and ↓ ↓ ↓ ↓ ), select the Continuity function
and press VAL to confirm. It is possible to enter the Function field
using key F2 and to select the function with the navigation keys (↑ ↑ ↑ ↑
and ↓ ↓ ↓ ↓ ).
Confirming the Continuity function calls up the following screen:
36/47