F. C
ALIBRATION AND
In the context of metrological quality monitoring, the user may have to carry
out a periodic performance verification.
The verification must take the standard metrological precautions into
consideration. The following instructions are to be applied.
The operations are carried out under reference conditions, namely:
•
Room temperature: 23°C ± 5°C.
•
Relative humidity: 45% to 75%.
The standards that constitute the measuring chain must be such that the
errors at the check points are known and are less than or equal to ± 0.008%.
It this verification reveals one or more characteristics of the instrument to be
outside the tolerances specified in the technical specifications chapter, you
can:
•
Either carry out the adjustment procedure given in the maintenance
document, which requires an instrument whose performance is at
least as good at that used for the preceding verification.
•
Or return the instrument to the address indicated below for
verification and calibration.
A
DJUSTMENT
AOIP SAS
Rue Dupont Gravé
F-14600 Honfleur
From France :
01.69.02.89.30
From your country :
+33(1) 69.02.89.50
Fax : +33(1) 69 02 89 60
Email : sav@aoip.com
43/47