Cleaning And Maintenance; Manual Cleaning - BD EnCor Adaptateur vertical GE Mode D'emploi

Système de biopsie mammaire
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 9
Table 1. Needle Parameters
A Vertical
B Notch
C End
Caution: Refer to the GE™ Senographe DS Instructions for Use
to place a Needle Guide Insert into the GE™ Vertical Needle
Guide Holder using an EnCor™ Reusable Bushing (GE UPNGB).

Cleaning and Maintenance

Note: The EnCor™ GE™ Vertical Adapter is not intended
to be sterilized. For Cleaning Instructions for the EnCor™
Fire Forward Accessory, refer to the EnCor™ Fire Forward
Accessory Instructions for Use.

Manual Cleaning

After each use, immediately clean the EnCor™ GE™ Vertical
Adapter with the steps outlined below.
1. Remove the EnCor™ Fire Forward Accessory (A) from the
EnCor™ GE™ Vertical Adapter (L) using the Thumb Screw
(I). Reference Figure 2.
2. Thoroughly wipe all external surface areas of the EnCor™
GE™ Vertical Adapter (L) with a cleaning wipe (for example,
CaviWipes1 or equivalent cleaning wipe).
3. For the steps mentioned above, ensure all surfaces of the
EnCor™ GE™ Vertical Adapter (L) are wiped continuously
and remain wet with a contact time of 3 minutes minimum.
4. Visually inspect the EnCor™ GE™ Vertical Adapter (L) to
in a well-lit area that has good lighting conditions (e.g
residual contaminants are not present.
Full Sample 7G, 10G, and
12G Probes (mm)
120.5
19.0
10.0
Half Sample 7G, 10G, and
12G Probes (mm)
6
120.5
10.0
10.0

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Ecgevertffadp

Table des Matières