If the device comes in contact with bodily fluid, the following manual cleaning instructions can be used:
1� Using enzymatic/neutral pH detergent, prepare per the manufacturer's instructions utilizing tap water entailing a temperature range of 27°C to 44°C
(81°F to 111°F)�
2� Do not submerge or spray the device with cleaning solution�
3� Use a soft, clean cloth saturated with the cleaning solution (not dripping) to manually clean all external contaminated surfaces of the device for a
minimum of 5 minutes� Be careful not to cause any liquid to drip into the charger or vacuum� In conjunction with this step, use an appropriately sized
brush (standard soft bristle toothbrush and lumen brush) to remove all visible soil from the device; paying particular attention to cracks, crevices and
other hard to clean areas� Note: It is recommended that enzymatic or neutral pH detergent solution should be changed when it becomes grossly
contaminated (bloody and/or turbid)�
4� Rinse thoroughly by using a soft, clean cloth saturated with lukewarm (27°C - 44°C) tap water (not dripping) for a minimum of 30 seconds� Repeat 2
additional times for a total of 3 total rinse cycles� Be careful not to cause any liquid to drip into the charger or vacuum�
5� Dry the device with a clean, lint-free towel�
6� Visually examine the device for cleanliness�
7� If visible soil remains, repeat cleaning procedure until the device is thoroughly clean�

Inspection/Maintenance

• The BD ClipVac™ Vacuum Unit requires no routine maintenance aside from cleaning system after each patient use� If a replacement is needed,
contact your local BD representative or customer service� Service or attempted repair performed by unqualified personnel may result in a risk of injury,
electric shock, fire, or permanent damage to the equipment and will void any warranty�
• Never use BD ClipVac™ Vacuum Unit or a battery that appears to be defective, damaged, or not working properly�
Limited Warranty:
• The BD ClipVac™ Presurgical Hair Removal System from BD (REFs 5500E, 5505, 5506A/5506E/5506U/5506K) is warranted against defects in material
and workmanship under normal use for a period of two (2) years from the date of unit placement� If this product is determined to be defective with
the warranty period, it will be repaired or replaced without charge by BD, if you contact your local representative or customer service representative�
They will provide the proper return authorization� Please include a description of the defect or damage with the returned unit�
• This warranty does not cover the Disposable Filter Assembly (REF 5575) nor does it cover damage caused by misuse, negligence, accident, or abuse,
alteration or modification of the unit by anyone other than BD or its authorized agent, or affixation of any accessory not manufactured or authorized
by BD, or used in combination with other surgical clippers�
• BD is not liable in either tort or contract for any cost, expense, loss, or damage whether direct, consequential, special, collateral, or incidental arising
out of the use, misuse, or inability to use this product� Except as expressly provided above, BD disclaims all warranties including, but not limited to the
implied warranties of merchantability and fitness for particular purpose�