EDS 300
9.2 SUMMARY OF THE BASIC SETTINGS
Setting
Switching mode switching output 1 (Sm 1)
Switching output 1 is operating in
switching point / hysteresis function
Switching output 1 is operating in
window function
Switching direction switching output 1 (S 1)
"ON ":
normally open function.
"OFF ":
normally closed function.
Switch-on delay switching output 1
(T
1)
on
Time in seconds which must elapse, once the
particular switching point has been reached or
exceeded, before switching will occur.
Switch-off delay switching output 1
(T
1)
off
Time in seconds which must elapse,
once the pressure has fallen below the
particular switch-back point, before
switching will occur.
Switching output 2, see above
Primary display (Primary)
Display value which should remain
permanently displayed:
"ACT .":
actual pressure
"Top":
pressure peak value
"S.P.1" or "S.P.2": switching point 1 or 2
"OFF ":
display off
(for function, see point 6 "Digital display")
Calibration of sensor zero point (Calibrate)
"YES":
The base pressure is saved as the new zero
point. This is possible in the range +/- 3 % of the
unit's nominal pressure range. "new" appears in
the display when a calibration is carried out in the
permissible range, otherwise "Err" is displayed.
This function is useful for example if there
is always a residual pressure in the system
which should however be displayed as 0 bar.
Important:
• Following a zero point adjustment, for example, on a 600 bar unit, a pressure of up to
18 bar is displayed as 0 bar. Before any work is carried out on the hydraulic system,
ensure that the system is de-pressurised. (for reference only)
Version number (Version)
Display of the current software
version. (for reference only)
To close basic settings (End)