Yearly Service
Inspection of Guard
34
Discotom-60 /-65
Instruction Manual
Visually inspect the guard and the glass for signs of wear or
damage.
Carry out inspection at more regular intervals if Discotom
is used for more than one 7 hour shift a day.
Struers recommends that the PETG glass is replaced after 5 years of
routine operation.
The guard should be replaced immediately if it has been weakened
by collision with projectile objects or if there are visible signs of
deterioration as a result of using a cooling fluid other than those
produced by Struers.
A label on the guard indicates when the glass is due to be replaced.
Important