Message/Error
Frequency inverter
communication error!
Contactor K1-K3 not
activated,
Please check the hood
Cutting motor blocked!
Please reduce the motor
load
Cutting motor will not
stop!
Please call service
technician
Frequency inverter fault!
Lenze error code: XX
Feed motor not
stopped!
Cutting table position
not found!
50
Discotom-60 /-65
Instruction Manual
Type
Explanation
ERROR
Data communication with
(#20)
the frequency inverter
failed (D-65 only).
The start button was
MESSAGE
(#21)
pressed, but none of the 3
motor contacts were
activated (D-60 only).
The cutting chamber
hood may not be fully
closed.
MESSAGE
During cutting, motor load
exceeded its safe limit.
(#22)
ERROR
A motor terminal failure
(#23)
occurred. After pressing
the stop button , the
motor did not stop.
(D-60 only).
ERROR
Frequency inverter failure
(#25)
(D-65 only).
ERROR
After pressing stop button
, the motor did not stop.
(#26)
ERROR
During the movement of
(#27)
Y-table, a position was
not found within a set
amount of time. The feed
motor module is most
likely defective.
Action
Switch off and on at Main Switch.
If the error message is still displayed,
contact a Struers service technician.
Check that the hood is closed properly.
If the error message is still displayed,
contact a Struers service technician.
Reduce feed speed or cutting force.
Check also that the sample is properly
secured as a sample that is not correctly
secured can trap the cut-off wheel with
enough force to stop it rotating.
Switch off the Discotom at the Main Switch
and call a Struers service technician.
Switch off the Discotom at the Main Switch
and call a Struers service technician.
Switch off the Discotom at the Main Switch
and call a Struers service technician.
Switch off the Discotom at the Main Switch
and call a Struers service technician.