Discotom-65 is factory mounted with an electric cable and an
additional wire for earth connection. Both MUST be hard-wired into
the mains (ref. EN 50178 / 5.2.11.1):
EU cable
L
Brown
1
L
Black
2
L
Black or grey
3
Earth
Yellow/green
Neutral Blue (Not used)
 The second Protective Earth must be earthed by connection to a
terminal marked
Alternatively, use a system whereby there is an automatic
disconnection of the supply in case of loss continuity of the
protective conductor.
For Electrical Installations with Residual Current Circuit Breakers
For Discotom-65 connected to electrical installations with residual current
circuit breakers, a residual current circuit breaker,
type B time delayed, 30 mA is REQUIRED
For Electrical Installations without Residual Current Circuit Breakers
The equipment must be protected by an insulation transformer
Please contact a qualified electrician to verify which option is suitable for the
Both requirements refer to the European standard EN 50178 / 5.2.11.1.
Similar standards apply in North America.
L
1
L
2
L
3
Earth
Neutral White (Not used)
or PE.
Important:
(ref. EN 50178 / 5.2.11.1).
(double-wound transformer)
local installation setup.
UL cable
Black
Red
Orange/ turquoise
Green (or Yellow/green)
Protective earth

Electric cable