SEE REQUIREMENTS FOR MAINS CABLE
- note #1
MUST BE EXTERNALLY FUSED
Fmax: See note #2
W1 EU
X1:L3
X1:L3
L3
L3
BK3
BK or GY
L3
X1:L2
X1:L2
L2
L2
BK
BK2
L2
X1:L1
X1:L1
L1
L1
BN
BK1
L1
X1:N
X1:N
QS1
QS1
BU
No internal
MAINS SWITCH
N
connection
GNYE
XPE 1
XPE 1
PE
CHASSIS
CHASSIS
4(5) x 2,5mm²
OR
W1 UL/CSA
X1:L3
X1:L3
OG
BK3
L3
QS1
X1:L2
X1:L2
RD
BK2
L2
MAINS SWITCH
X1:L1
X1:L1
BK
BK1
L1
X1:N
X1:N
WH
No internal
N
connection
GN (GNYE)
PE
XPE 1
XPE 1
CHASSIS
CHASSIS
4(5) x AWG 12
VOTAGE / FREQ.
W1 - mains cable
3 x 200-210V / 50-60Hz
AWG = 12
3 x 220-230V / 50Hz
2.5mm^2
3 x 220-240V / 60Hz
AWG = 12
3 x 380-415V / 50Hz
2.5mm^2
AWG = 12
3 x 460-480V 60Hz
T3
T3
BK13
T2
T2
BK12
T1
T1
BK11
See table
A10
BK7
Powersupply 240W
L1
+
SMPS
+
24V DC / 10A
BK8
L2
-
-
200-500Vac
GNYE4
PE
Input
XPE 3
For connection, see also page 3
F1+F2
note #1
note #2
(fuse size)
F3
T2
max. ext. fuse
(fuse size)
Number of wires
See note #3
3 x 40AT
2 x 4AT (CC)
5AT
2
3 x 40AT
2 x 4AT (aM)
5AT
2
3 x 40AT
2 x 4AT (CC)
5AT
2
3 x 40AT
2 x 2AT (aM)
5AT
2
3 x 40AT
2 x 2AT (CC)
5AT
2
A2
Brake module
F3
F3
RD3
RD4
5,0 AT FUSE
5,0 AT FUSE
R1
BK29
BK28
V1
RD5
BK27
RD1
RD2
K5
Note #3:
F1, F2, F3, F5 are time delay fuses
CC...Class-CC characteristic
aM...aM characteristic
All wires AWG 12
BK4
AWG14
BK5
AWG14
1
3
5
1
3
5
K1
2
4
6
2
4
6
T2 (current transformer)
1
3
5
K2
STAR
2
4
6
XPE 2
U1
V1
W1
PE
MAIN MOTOR
Rev B. 27/07/07 TKO:Trafo T1
REV E: SPE 2009.05.07
replaced by SMPS A10,
- See page 2
REV F: AKR 2009.11.26
correction on diagram, variant table
- X1:N added
changed, X3 corrected to X1,
- A10 changed to variant part.
GNYE4 changed to GNYE6
- LED lighting added
- Embedded SMU added
REV G: AKR 2012.08.17
Rev C: FTH 13-05-2008
- A10 changed so it's no longer
Connections on M1 changed
variant part
Rev D: FTH 21-05-2008
New aux.contacts on K1 to K5
Friday, August 17, 2012
Friday, August 17, 2012
Friday, August 17, 2012
DELTA
BK6
1
3
5
1
3
5
K3
K4
2
4
6
2
4
6
c.w. rotation
U2
M1
W2
3 x U
Y / D
V2
100R @ 0C
or
300R @ 0C
(System will autodetect)
Sruers A/S
Sruers A/S
Sruers A/S
Pederstrupvej 84
Pederstrupvej 84
Pederstrupvej 84
DK-2750 Ballerup
DK-2750 Ballerup
DK-2750 Ballerup
Denmark
Denmark
Denmark
telephone: +45 44 600 800
telephone: +45 44 600 800
telephone: +45 44 600 800
DISCOTOM-60, MAIN SUPPLY CIRCUIT
DISCOTOM-60, MAIN SUPPLY CIRCUIT
DISCOTOM-60, MAIN SUPPLY CIRCUIT
Size
Size
Size
CAGE Code
CAGE Code
CAGE Code
DWG NO
DWG NO
DWG NO
Rev
Rev
Rev
15903100
15903100
15903100
G
G
G
A2
A2
A2
Scale
Scale
Scale
Sheet
Sheet
Sheet
1
1
1
of
of
of
3
3
3