Replacing The Filter (For Instruments With Integral Pump); Cleaning The Instrument; Storage; Shipment - MSA ALTAIR 5 Mode D'emploi

Masquer les pouces Voir aussi pour ALTAIR 5:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

MSA
Calibration is required after a sensor is installed; otherwise, the
instrument will not perform as expected and persons relying on this
product for their safety could sustain serious personal injury or death.
4.5.

Replacing the Filter (for instruments with integral pump)

(1) Turn OFF the instrument.
(2) Unscrew the two captive screws from the clear filter cover on the back
of the instrument to access the filter.
(3) Carefully lift out the O-ring and the filter disk.
(4) Place the new filter disk in the recess.
(5) Replace the O-ring in the recess.
(6) If the instrument is used in dusty environments, the fibrous dust filters
supplied in Maintenance Kit 10094829 may be used in addition to the
filter disk. Place the fibrous dust filter in the clear filter cover.
If the instrument is configured for a reactive toxic gas (Cl2, ClO2, NH3),
use only the filters supplied in the Reactive Gas Maintenance Kit
10099533.
(7) Re-install the clear filter cover on the back of the instrument.
4.6.

Cleaning the Instrument

Clean the exterior of the instrument regularly using only a damp cloth.
Do not use cleaning agents.
4.7.

Storage

When not in use, store the instrument in a safe, dry place between 65°F
and 86°F (18 °C and 30 °C). After storage, always recheck instrument
calibration before use. If not to be used in 30 days remove battery pack.
4.8.

Shipment

Pack the instrument in its original shipping container with suitable padding.
If the original container is unavailable, an equivalent container may be
substituted.
US
WARNING
CAUTION
ALTAIR 5 - Operating Manual
MAINTENANCE
49