OPERATION
3.4.1 Calibration Options
Monochrome display
CALIBRATION OPTIONS
The Calibration Options menu allows the user to:
modify the calibration cylinder settings
(
CYLINDER SETUP
enable/disable calibration due and to set the number of days
(
CAL DUE OPTIONS
enable/disable the option to show the last cal date at turn on and
(
LAST CAL DATE
enable/disable the option for password protected calibration
(
CAL PASSWORD
Press
the
the
the
Setting Calibration Cylinder
This option has a dialog similar to the span calibration dialog.
The display shows all active sensors.
(1) Press the
The screen for the first calibration cylinder displays.
(2) Press
the
the
With this confirmation the instrument automatically moves to the next
cylinder setting.
(3) Repeat the sequence for changing the required settings for all
necessary gas values.
After the last setting is performed, the instrument returns to the
Calibration Options menu.
30
)
)
)
)
button go to next page
button to go previous page
button to enter setup.
button to enter setup.
or
button to change the value.
button to confirm the setup.
ALTAIR 5 - Operating Manual