8.4. ATC - Absolute Temperature Calibration

Internal sensor (T
F
Measuring point(T
M
Measuring point
72
ATC serves to compensate a temperature difference that might
occur between circulator and a defined measuring point in the bath
tank because of physical properties.
)
The difference temperature is determined (∆T= T
and stored as correcting factor (example ∆T = -0.2 °C).
)
1.
Press the cursor key
2.
Use the cursor keys
„At0".
3.
With the edit keys
enter
Enter the corrective value.
4.
Using the cursor keys
press enter
5.
Press
The temperature on the measuring point rises to a temperature of
37.0 °C and is indicated on the
MULTI-DISPLAY (LED).
The ATC function stays activated until resetting to 00.0 °C.
Recommendation:
Use a calibrated temperature measuring instrument.
.
set the correcting factor (example -0.20 °C) and then
.
and
at the same time.