Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Preparations

5.3. Filling / Draining

Warning:
There are thermal dangers when opening the bath cover:
Burn, scald, hot steam, hot parts and surfaces that can be touched.
 Do not drain the bath fluid while it is hot!
 Check the temperature of the bath fluid prior to draining
62
Filling:
• Maximum filling level: 6 cm below the bath rim
• Minimum filling level: 7 cm (level approx. 3 cm above
surface (16) for the placement of items/fixtures)
Note:
The working filling level depends on size and number of the
items (fixtures) to be placed inside.
Fill to minimum level only. Insert the complemented shaking
carriage and correct the filling level (adding or removing
liquid) as required
Use the water bath cover to keep temperature losses to a
minimum:
Lift-up Makrolon
Draining:
• Switch off the shaking water bath with the mains switch
and move the equipment to the table edge.
• Place a suitable collecting bucket or tub underneath the
equipment for draining the used liquid.
• To drain the liquid open the dainage screw (11) on the side
of the water bath.
• After the liquid has been fully drained, securely tighten the
drainage screw (11) again.
(by switching the unit on for a short moment, for example).
cover see page 64

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Sw23

Table des Matières