Cleaning / Repairing The Unit - Julabo SW22 Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

13. Cleaning / repairing the unit

Caution:
Improper maintenance or repair can result in electric shock or damage to the
unit.
 Repairs and any other work are to be carried out only by qualified service
personnel authorized by JULABO.
 Always turn off the unit and disconnect the mains cable from the power
source before performing any service or maintenance procedures, or
before moving the unit.
 Prevent humidity from entering into the water bath.
 Do not use alcohol-based or solvent-based cleaning agents.
These cleaning agents will result in damage and cracks in the Makrolon
cover.
Cleaning:
Repairs:
Returning a unit:
For cleaning the bath tank and the immersed parts of the water
bath, use low surface tension water (e.g., soap suds).
Clean the outside of the unit using a wet cloth and low surface
tension water.
The JULABO Skaking Water Baths are designed for continuous
operation under normal conditions. Periodic maintenance is not
required.
The tank should be filled only with a bath fluid recommended by
JULABO. To avoid contamination, it is essential to change the bath
fluid from time to time.
Before asking for a service technician or returning a JULABO
instrument for repair, please contact an authorized JULABO
service station.
When returning the unit:
• Clean the unit and, if necessary, decontaminate the unit in order
to avoid endangering service personnel.
• Attach a short fault description.
• During transport the unit has to stand upright. Mark the packing
correspondingly.
• When returning a unit, take care of careful and adequate
packing.
JULABO is not responsible for damages that might occur from
insufficient packing.
JULABO reserves the right to carry out technical modifications with
repairs for providing improved performance of a unit.
Shaking Water Baths