Troubleshooting Guide / Error Messages - Julabo SW22 Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Troubleshooting guide / Error messages

10. Troubleshooting guide / Error messages
))))
+
60
45
4 s
30
76
Whenever the microprocessor electronics registers a failure, a
complete shutdown of the heater and circulation pump is performed.
The alarm light "
sounds.
Cause:
The waterbath is operated without bath liquid, or the liquid level
is insufficient
or
The adjusted temperature limit was exceeded or the temperature
fell below the limit.
Remedy: Replenish the bath tank with the bath liquid.
Control the adjustment of the temperature limit.
Get to safety the samples.
The wires of the working temperature sensor are interrupted or
short-circuited.
other errors
After eliminating the malfunction, press the mains power switch off
and on again to cancel the alarm state.
If the unit cannot be returned to operation, contact an authorized
15
JULABO service station.
" illuminates and a continuous signal tone
• Heating circuit interrupted.
• Short-circuit of triac.
• Short-circuit in alarm relay.

Hide quick links:

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Sw23

Table des Matières