Julabo SW22 Mode D'emploi page 57

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Safety installations according to IEC 61010-2-010:
Excess temperature protection
Classification according to DIN 12876-1
Alarm indication
Supplementary safety installations
High temperature warning function
Low temperature warning function
Timer
Environmental conditions according to EN 61 010, part 1:
Use only indoor.
Altitude up to 2000 m - normal zero.
Ambient temperature: +5 ... +40 °C (for storage and transportation)
Air humidity:
Max. rel. humidity 80 % for temperatures up to +31 °C,
linear decrease down to 50 % relative humidity at a temperature of +40 °C
Max. mains fluctuation of ±10 % are permissible.
Protection class according to EN 60 529
The unit corresponds to Class I
Overvoltage category II
Pollution degree 2
Caution:
The unit is not for use in explosive environment
Standards for interference resistance according to EN 61326-1
This unit is an ISM device classified in Group 1 (using high frequency for internal purposes)
Class A (industrial and commercial range).
105 °C - fixed value
class I
optical + audible (continuous tone)
optical + audible (in intervals)
optical + audible (in intervals)
audible (in intervals)
IP 21
Shaking Water Baths
55

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Sw23

Table des Matières