Preparations
6.2

Bath fluids

Pos: 13 /T echnische Dokum entation/Flash CORIO/PURA/Befüllen-Entleeren @ 1 \mod_14 576 034 970 29_ 18.docx @ 940 1 @ 2 @ 1
6.3

Filling / Draining

Recommended bath fluids: soft/decalcified water.
Poor water quality may result in corrosion in the bath.
The quality of water (tap water) depends on local conditions.
Ferrous water can cause corrosion - even on stainless steel.
Chloric water can cause pitting corrosion.
Distilled and deionized water is unsuitable. Their special properties
cause corrosion in the bath, even in stainless steel.
Due to the high concentration of lime, hard water is not suitable
for temperature control because it leads to calcification in the bath.
JULABO takes no responsibility for damages caused by the selection of
an unsuitable bath fluid.
Please contact JULABO before using other than recommended bath
fluids.
Do not use flammable bath fluids!
Water bath protection products
The water bath protection product "Aqua-Stabil" is recommended to
combat algae, bacteria and fungus formation.
Order No.
8 940 006
8 940 012
Thermal dangers when opening the bath cover:
Burn, scald, hot steam, hot parts and surfaces that can be touched.
Do not drain the bath fluid while it is hot!
Check the temperature of the bath fluid prior to draining, by
switching the unit on for a short moment, for example.
Filling:
Recommended maximum filling level is 2.5 cm below the tank
rim.
Minimum filling level: approx. 2 cm above the bath bottom.
Designation
6x 100 ml bottles
12x 100 ml bottles