13. EMC / Technical description
The Freestyle breastpump needs special precautions regarding EMC (Electromag-
netic compatibility) and needs to be installed and put into service according to the
EMC information provided in these instructions for use. Wireless communications
equipment such as wireless home network devices, mobile phones, cordless
telephones and their base stations, walkie-talkies can affect the electric breastpump
Freestyle and should be kept at least a distance 1.0 m away from the equipment.
Electromagnetic Compatibility (EMC, IEC 60601-1-2:2007, Table 1)
Electromagnetic emissions
The electric breastpump Freestyle is intended for use in the electromagnetic environment
specified below. The customer or the user of the electric breastpump Freestyle should
assure that it is used in such an environment.
Emission tests
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
No significant voltage changes
a
184
Warning – The electric breastpump Freestyle should not be used adjacent
184
to or stacked with other equipment. If adjacent or stacked use is necessary,
the electric breastpump Freestyle should be observed to verify normal operation
in the configuration in which it will be used.
Electromagnetic environment –
Compliance
guidance
The electric breastpump Freestyle uses RF
energy only for its internal function. Therefore,
Group 1
its RF emissions are very low and are not likely
to cause any interference in nearby electronic
equipment.
Class B
The electric breastpump Freestyle is suitable
for use in all establishments, including domestic
Class A
establishments and those directly connected
to the public low-voltage power supply network
that supplies buildings used for domestic pur-
Not applica-
poses.
ble
a