EMC / Technical Description

Electromagnetic Compatibility (EMC, IEC 60601-1-2:2001)
IEC 60601-1-2, Table 201

Electromagnetic Emissions

The Symphony breastpump is intended for use in the electromagnetic environment specified below.
The customer or the user of the Symphony breastpump should assure that it is used in such environment.
Emission Tests
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations /
flicker emissions
IEC 60000-3-3
Warning – The use of accessories and cables other than those specified may result in increased
emissions or decreased immunity of the Symphony breastpump.
The Symphony breastpump should not be used adjacent to or stacked with other equipment.
If adjacent or stacked use is necessary, the Symphony breastpump should be observed to verify
normal operation in the configuration in which it will be used.
The Symphony breastpump was found to have no essential performance by a risk analysis and thus,
was not tested for immunity to electromagnetic disturbances (IEC60601-1-2:2001, 6.8.3.201 i)
82
Compliance
Electromagnetic environment - guidance
The Symphony breastpump uses RF energy only for its
Group 1
internal function. Therefore, its RF emissions are very
low and are not likely to cause any interference in nearby
electronic equipment.
Class B
The Symphony breastpump is suitable for use in all
establishments, including domestic establishments
Class A
and those directly connected to the public low-voltage
power supply network that supplies buildings used for
domestic purposes.
Complies

Publicité