13. EMC / Technical description

The Mini Electric breastpump needs special precautions regarding EMC (Electro-
magnetic compatibility) and needs to be installed and put into service according to
the EMC information provided in these instructions for use. Wireless communica-
tions equipment such as wireless home network devices, mobile phones, cordless
telephones and their base stations, walkie-talkies can affect the electric breastpump
Mini Electric and should be kept at least a distance 1.0 m away from the equipment.
Electromagnetic Compatibility (EMC, IEC 60601-1-2:2007, Table 1)
Electromagnetic emissions
The electric breastpump Mini Electric is intended for use in the electromagnetic environment
specified below. The customer or the user of the electric breastpump Mini Electric should
assure that it is used in such an environment.
Emission tests
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
a
No significant voltage changes
160
Warning – The electric breastpump Mini Electric should not be used adjacent to or
stacked with other equipment. If adjacent or stacked use is necessary, the electric
breastpump Mini Electric should be observed to verify normal operation in the configurati-
on in which it will be used.
Electromagnetic environment –
Compliance
guidance
The electric breastpump Mini Electric uses RF
energy only for its internal function. Therefore,
Group 1
its RF emissions are very low and are not likely
to cause any interference in nearby electronic
equipment.
Class B
The electric breastpump Mini Electric is suitable
Class A
for use in all establishments, including domestic
establishments and those directly connected to
the public low-voltage power supply network that
supplies buildings used for domestic purposes.
Not
applicable
a