7. EMC / Technical description
The Swing breastpump needs special precautions regarding EMC (Electromagnetic
compatibility) and needs to be installed and put into service according to the EMC
information provided in these instructions for use. Wireless communications
equipment such as wireless home network devices, mobile phones, cordless
telephones and their base stations, walkie-talkies can affect the electric breastpump
Swing and should be kept at least a distance 1.0 m away from the equipment.
Electromagnetic Compatibility (EMC, IEC 60601-1-2:2007, Table 1)
Electromagnetic emissions
The electric breastpump Swing is intended for use in the electromagnetic environment specified below.
The customer or the user of the electric breastpump Swing should assure that it is used in such an
environment.
Emission tests
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions IEC
61000-3-2
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
a
No significant voltage changes
Warning
Warning
The electric breastpump Swing should not be used adjacent to or stacked with other
equipment. If adjacent or stacked use is necessary, the electric breastpump Swing
should be observed to verify normal operation in the configuration in which it will be used.
Questions? Visit www.medela.com or call us at 1-800-435-8316.
Compliance
Electromagnetic environment – guidance
The electric breastpump Freestyle uses RF
energy only for its internal function. Therefore,
Group 1
its RF emissions are very low and are not likely
to cause any interference in nearby electronic
equipment.
Class B
The electric breastpump Freestyle is suitable
for use in all establishments, including domestic
Class A
establishments and those directly connected
to the public low-voltage power supply network
that supplies buildings used for domestic
Not applicable
a
purposes.
31