Emc / Technical Description; Electromagnetic Emissions - Medela Swing Mode D'emploi

Masquer les pouces Voir aussi pour Swing:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

EMC / Technical Description

Electromagnetic Compatibility (EMC, IEC 60601-1-2:2001)
IEC 60601-1-2, Table 201

Electromagnetic Emissions

The Swing is intended for use in the electromagnetic environment specified below.
The customer or the user of the Swing should assure that it is used in such environment.
Emission Tests
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations /
flicker emissions
IEC 60000-3-3
Warning – The use of accessories and cables other than those specified may result in increased emissions or
decreased immunity of the Swing breastpump.
The Swing breastpump should not be used adjacent to or stacked with other equipment.
If adjacent or stacked use is necessary, the Swing breastpump should be observed to verify normal operation in the
configuration in which it will be used.
The Swing breastpump was found to have no essential performance by a risk analysis and thus, was not tested for
immunity to electro magnetic disturbances (IEC60601-1-2:2001, 6.8.3.201 i)
48
Compliance
Electromagnetic environment - guidance
The Swing uses RF energy only for its internal function. Therefore,
Group 1
its RF emissions are very low and are not likely to cause any inter-
ference in nearby electronic equipment.
Class B
The Swing is suitable for use in all establishments, including
domestic establishments and those directly connected
Class A
to the public low-voltage power supply network that supplies
buildings used for domestic purposes.
Complies

Publicité

Table des Matières
loading

Table des Matières