Full HD 3CCD Camera system PV440
►
To prevent damage caused by improper setup or operation, and in order
not to compromise warranty and manufacturer liability:
– Only use the product according to these instructions for use.
– Follow the safety and maintenance instructions.
– Only combine Aesculap products with each other.
►
Ensure that the product and its accessories are operated and used only
by persons with the requisite training, knowledge, or experience.
►
Keep the instructions for use accessible for the user.
►
Always adhere to applicable standards.
►
Do not operate the product in the vicinity of flammable anesthetics.
►
Check the power cord for leakage currents and correct grounding at
regular intervals.
3.
Product description
3.1
Scope of supply
Full HD 3CCD Camera system PV440, consisting of:
■
Full HD camera control unit
■
Full HD 3CCD Camera head
■
DVI-D cable
■
Instructions for use
3.2
Components required for operation
In addition to the camera system, the following components are required
for an endoscopic intervention:
■
Endo-lens
■
Power cord
■
insert the endoscope
■
Light source
■
Light cable
■
Monitor
3.3
Intended use
The Aesculap Full HD (=High Definition) camera is an endoscopic video
camera to transmit real-time video images of a rigid or flexible endoscope,
via the use of a lens with endocoupler, to a video monitor or other related
documentation devices.
The Aesculap Full HD camera is suitable for all endoscopic applications.
The current use of the camera involves connecting a lens with endocou-
pler to the camera. An endoscope with standard eyepiece is then con-
nected to the endocoupler and held in place by the grasping mechanism
of the endocoupler.
The Full HD camera is the preferred camera for endoscopic procedures due
to its superior image quality of 1920 x 1080 pixel and progressive scan
mode.
4
®
3.4
Operating principle
An endoscope visualizes the operation site and an objective lens focuses
this image onto the three CCD sensors, where it is converted into digital
signals. The image brightness is automatically adjusted according to the
light intensity (autoshutter).
Through camera cable 4, the digital image signals are transferred to cam-
era control unit 7, where the digital data are processed to be displayed on
the monitor.
The video signal processed in this way is then picked up at the video out-
put monitor, and digital documentation devices.
The camera system offers automatic, electronic brightness enhancement
for areas that are particularly difficult to illuminate, as well as five differ-
ent fixed mode settings and three user adjustable mode settings for vari-
ous situations during operation.
The video camera system is also equipped with a white adjustment system
for optimal color representation.
The lens and an additional electronic zoom feature ensures that the image
on the monitor is displayed at the optimum size.
4.
Preparation and setup
Non-compliance with the following instructions will preclude all respon-
sibility and liability in this respect on the part of Aesculap.
►
When setting up and operating the product, adhere to
– national regulations for installation and operation,
– national regulations on fire and explosion protection.
Note
For the safety of patients and users it is essential that the mains power cord
and, especially, the protective earth connection are intact. In many cases
defective or missing protective earth connections are not registered imme-
diately.
►
Connect the device via the potential equalization terminal at the rear
panel of the device to the potential equalization system of the room
used for medical purposes.
Note
The potential equalization lead can be ordered from the manufacturer as
art. no. GK535 (4 m length) or TA008205 (0.8 m length).
4.1
First use
Risk of injury and/or product malfunction due to
incorrect operation of the electromedical system!
►
Adhere to the instructions for use of any medi-
WARNING
cal device.