| INSTRUCTIONS FOR USE
GB

INDICATION

The paediGAV is intended for use in cases of
pediatric hydrocephalus for the drainage of
cerebrospinal fluid from the ventricles into the
peritoneum.

TECHNICAL DESCRIPTION

The paediGAV is a position-sensitive valve for
the treatment of hydrocephalus. It comprises a
ball-cone unit and a gravitational unit. By this de-
sign, physiological drainage can be assured in-
dependent of the physical position of the patient.
gravitational unit
14
Fig. 1 shows a schematic cross section of
the paediGAV. It consists of a sturdy titanium
casing, which houses, in its proximal section,
a welltried ball-cone unit. A spiral spring (2)
maintains the opening pressure of the ball-cone
unit. The gravitational unit in the distal part con-
sists of a tantalum ball (3), which defines the
opening pressure of this valve, and a sapphire
ball (4), which ensures its precise closure.
As an option, a connector or a silicone catheter
is installed at the distal end of the valve. Coding
rings (1) allow the identification of the pressure
setting in an x-ray image. The paediGAV regu-
lates the intraventricular pressure (IVP) for the
patient being in a horizontal or in a standing
position.
ball-cone unit
1 coding ring
2 spiral spring
3 tantalum ball
4 sapphire ball
Fig. 1: Schematic cross section of the paediGAV
paediGAV
1
2
3
4