Vercise™ DBS Programming Manual
The two rates are dependent on one another in that only certain combinations of rates are
permitted. The table below indicates which frequencies are available for the second Area
at commonly used frequencies for the first Area.
First Area
Rate (Hz)
99
132
179
238
*Subset includes only frequencies in typical DBS range (89 to 255 Hz) using typical pulse
widths (30 to 90 μs).
†Available when both pulse widths are less than or equal to 60 μs.
Available when both pulse widths are less than or equal to 30 μs.
‡
Vercise™ DBS Programming Manual
91053785-17
30 of 1705
Subset of Available Second Area Rates (Hz)*
89
, 99, 112
, 113
, 116
†
†
†
159
, 170
, 174
, 179
†
‡
†
†
99, 113
, 116
, 119
, 132, 149
†
‡
‡
198, 231
, 238
‡
‡
89, 99
, 104
, 112
, 119, 128
†
‡
†
223
, 238
†
†
89
, 95, 99
, 119, 132
†
‡
198
, 238
‡
, 119, 128
, 132, 139
†
†
, 198, 223
, 231
, 238
†
†
, 154
, 159
, 170
‡
‡
†
, 139
, 143
, 149
†
‡
†
, 136
, 143
, 149
, 159, 170
‡
‡
†
‡
, 149, 154
,
†
‡
‡
, 176, 185
,
‡
‡
, 179, 208
,
†
‡
, 179
, 190
,
‡
†
‡