Les langues disponibles

16
MAINTENANCE SCHEDULE FOR THE OXFORD/HOYER ADVANCE
All Oxford/Hoyer products are designed for minimum maintenance, however some
safety checks and procedures are required. A schedule of DAILY tasks is detailed below.
A list of replacement parts is available upon request.
DAILY CHECK LIST
Sunrise Medical strongly recommends the following checks be carried out on a daily
basis and before using lift.
- MAKE sure the lift moves freely on its castors.
- MAKE sure the spreader bar is free to rotate and swing. Check the spreader bar is
firmly attached to the boom.
- EXAMINE the sling hooks on the spreader bar and side suspenders for excessive
wear. If in doubt - do not use.
- EXAMINE slings for fraying or other damage. DO NOT use any sling if damaged
- MAKE sure the legs open and close correctly.
- OPERATE the hand control or the hydraulic unit to confirm the boom raises and
lowers satisfactorily.
- CONFIRM the lift is not giving a low battery alarm when the hand control is
operated (electric lifts only). If the alarm sounds DO NOT use and place on
charge immediately.
- ON electric powered lifts check the operation of the emergency stop button.
- ON hydraulically operated lifts check for hydraulic fluid leakage. Any leakage
should be reported to a service engineer immediately and the lift should not be
used until it has been repaired.
MAINTENANCE, INSPECTION AND TEST
Sunrise Medical recommends a thorough inspection and test of the Oxford/Hoyer
Advance lift and lifting accessories; slings, etc. are carried out on a regular basis.
Inspection frequency varies from country to country, so you must check with your local
government agency how often an inspection is required. The examination and test
should be conducted according to the recommendations and procedures below.
Whenever possible Sunrise Medical recommends maintenance, inspection and certified
testing is carried out by authorised service agent / dealers only.
NOTE–
These recommendations are in compliance with the requirements of 1998 No2307 Health
and Safety: The Lifting Operations and Lifting Equipment Regulations 1998. This is a UK
regulation. Outside the UK please check your local country requirements.
294000.10003 Rev. B
E n g l i s h
E n g l i s h
*THESE CHECKS SHOULD INCLUDE:
1. SPREADER BAR: Check the spreader bar for freedom of rotation
and swing. Check for wear on the central pivot. Check for firm
attachment to the boom.
2. BOOM: Check the attachment of the boom to the mast. Make sure
there is only minimal side movement of the boom and the boom
is free to rotate on the boom bearing.
3. MAST: Check the operation of the mast-locking device. Make sure
the mast fully engages into the socket. Check the bottom actuator
or hydraulic unit mounting.
4. ACTUATOR (electric only): The actuator should require no
maintenance other than checking for correct operation and
listening for unusual noise.
5. POWER PACK (electric only): Check the function of the
emergency stop button and emergency down/up.
6. BATTERIES (electric only): The batteries are housed in the power
pack and should not require maintenance other than the regular
charging as detailed in the charging instructions. Check that the
connections remain clean.
7. HYDRAULIC UNIT The hydraulic unit should require no
maintenance other than checking for correct operation and
leakage of hydraulic fluid.
8. LEG ADJUSTMENT: Check the legs operate in both full
extensions (inward/outward).
9. CASTORS: Check all castors for firm attachment to the legs.
Check for free rotation of the castor and the wheels.
10. CLEANING: Clean with ordinary soap and water and/or any hard
surface disinfectant. Harsh chemical cleaners or abrasives should
be avoided as these may damage the surface finish of the lift.
Avoid wetting any of the electrical parts.
11. LOAD TEST: The load test should be carried out in accordance
with the manufacturer's test procedures.
12. BASE AND WHEELS: Ensure base is even and level
(all four wheels are on the floor).
13. SLINGS: Check for wear and fraying.
14. LUBRICATION: Oil pivot joints, including mast and boom
connections, pedal assembly, spreader bar joint (only if required
17






















).
294000.10003 Rev. B