E N G L I S H
3. Replace with a new filter. Additional filters should be purchased from your Sunrise provider.
4. Push new filter into position, and replace filter cover.
CAUTION– Do not reuse or wash filter or substitute any other material such as cotton for a Sunrise
air-inlet filter. It will result in compressor damage.
MAINTENANCE
All maintenance must be performed by a authorized Sunrise provider.
Electric shock hazard. Do not remove compressor cabinet. All disassembly and
maintenance must be done by a authorized Sunrise provider.
SPECIFICATIONS
Size ......................................................3.34" W x 2.2" H x 3.8" D (8.5 cm x 5.6 cm x 9.7 cm) w/o battery
Weight (approximate) .............................. 0.8 lbs. (0.36 kg) w/o battery or 1.5 lbs (0.68 kg) with battery
Electrical Requirements........................................................................100-240 V, 50-60 Hz; 12 VDC, 1.5A
Sound Level ........................................................................................................................................54 dBA
Power Consumption........................................................................................................25 watts maximum
Max. Compressor Pressure ............................................................................30 psig or greater* (206 kPa)
Compressor Free Air Flow ................................................................................................6 lpm or greater*
Intermittent Use ........................................................................................30 minutes On / 30 minutes Off
Operating Temperature Range ................................................................+40° to +104°F (+5° to +40°C)
Operating Humidity............................................................................................up to 95% non-condensing
Storage/Transport Temperature Range ....................................................-40° to +158° F (-40° to +70° C)
Storage/Transport Humidity ..............................................................................up to 95% non-condensing
Supplied Disposable Nebulizer Operating Pressure ....................................................8 psig (min) (55 kPa)
Supplied Reusable Nebulizer Operating Pressure ........................................................8 psig (min) (55 kPa)
Supplied Disposable Nebulizer Liter Flow ..................................................................................4 lpm (min)
Supplied Reusable Nebulizer Liter Flow......................................................................................4 lpm (min)
Supplied Disposable Nebulizer Capacity ........................................................................................6 ml (cc)
Supplied Reusable Nebulizer Capacity ............................................................................................6 ml (cc)
Supplied Disposable Nebulizer Nebulization Rate......................................................Approx. 0.20 ml/min*
Supplied Reusable Nebulizer Nebulization Rate ........................................................Approx. 0.20 ml/min*
Supplied Disposable Nebulizer MMAD ......................................................................................5 microns**
Supplied Reusable Nebulizer MMAD..........................................................................................5 microns**
*Conditions may vary based on altitude above sea level, barometric pressure and temperature.
**Value determined with minimum flow rate through the nebulizer (20°C, sea level).
Equipment Classifications
Protection against electric shock..........................................................Class II
Degree of protection against electric shock ......................................Type BF
Equipment not suitable for use in the presence of flammable anaesthetic mixture with air or with oxygen
or nitrous oxide.
DANGER
3.34" W x 2.2" H x 5.3" D (8.5 cm x 5.6 cm x 13.5 cm) with battery
9
9
S P E C I F I C AT I O N S
A-6910